vPad-XSC for Impulse 4000 USER MANUAL
xctrl “Setting SRP limit: max 300ms" | lim9,max,300
xtest “Wait! Then verify refractory test result" | PRFS
For the above “
xtest
” statement,
PRF
commands
Phase 3
to immediately begin testing the pacer refractory period, and
S
is a placeholder, which may also be “
L
”
Since the PRP limit and SRP limit are enabled prior to running the test, the Interface automatically
assigns a “PASS” result if the PRP is between 200 ms and 500 ms, and SRP is less than 300 ms.
For the case of a "PASS", the Test Report will contain an entry similar to the following:
Operation/Chapter 3
#
Page 21