vPad-XSC for Impulse 4000 USER MANUAL
Assuming that defib energy and sync delay limits are set to defaults, the Interface automatically
assigns a “PASS” result if the energy measured is 20 J ±15 % and sync delay is between 20 ms and
65 ms. For the case of a "PASS", the Test Report will contain an entry similar to the following:
___________________________________________________________
Task:
Set defib to sync mode and discharge now!
Data:
Defibrillator Sync Delay Test
Test:
Energy Limit: 20 J +/- 15 %
Energy: 20.2 J
Delay Limit: Min 20 ms Max 64 ms
Result: 32.2 ms
Result:
PASS
___________________________________________________________
Operation/Chapter 3
#
Page 18