S/5 Aespire
4-14
1009-0307-000
•
Put your hand on the inlet of the leak test device. Push hard for a
good seal.
•
If the bulb inflates in less than 60 seconds,
replace the leak test device.
4. Turn off all vaporizers.
5. Test the anesthesia machine for low-pressure leaks:
•
Turn the flow controls one and a half turns counterclockwise.
•
Connect the test device to the auxiliary gas outlet.
•
Compress and release the bulb until it is empty.
•
The vacuum causes the floats to move. This is usual. If the bulb
inflates in 30 seconds or less, there is a leak in the low-pressure
circuit. Refer to the “Pneumatic problems” table in “Alarms and
Troubleshooting” of Part 2 of this manual).
•
Disconnect the test device.
6. Test each vaporizer for low-pressure leaks:
•
Set the vaporizer to 1%.
•
Repeat step 5. If there is a low pressure leak, refer to Part 2 of this
manual.
7. Keep the test device with the system.
8. Turn all flow controls fully clockwise (minimum flow). Do not over tighten.
w
w
w
w
WARNING
Agent mixtures from the low-pressure leak test stay in the system.
Always flush the system with O
2
after the low-pressure leak test (1
L/min for one minute).
w
w
w
w
WARNING
Turn off all vaporizers at the end of the low-pressure leak test.
9. Flush the system with O
2
:
•
Set the system switch to on.
•
Set the O
2
flow to 1 L/min.
•
Continue the O
2
flow for one
minute.
•
Turn the O
2
flow control fully clockwise (minimum flow).
•
Set the system switch to Standby.