4 Preoperative Tests
1009-0307-000
4-13
•
If O
2
flow decreases more than 1 L/min, then install a different
vaporizer and try this step again.
•
If the O
2
flow decreases less than 1 L/min with a different vaporizer,
the malfunction is in the first vaporizer.
•
If the O
2
flow also decreases more than 1 L/min with a different
vaporizer, the malfunction is in the anesthesia system. Do not use
the system until it is serviced.
5. Complete steps 3 and 4 for each vaporizer.
6. Set the system switch to Standby.
Low-pressure leak test
w
w
w
w
WARNING
Do not use a system with a low-pressure leak. Anesthetic gas will
go into the atmosphere, not into the breathing circuit.
Perform either the Negative low-pressure leak test or the ISO 5358 or BSI
standard positive low-pressure leak test. It is not necessary to perform both
types of tests.
Negative low-pressure
leak test
1. Set the System switch to Standby.
2. Turn on the auxiliary common gas
outlet (switch in the ACGO
position).
3. Test the negative pressure leak test device. (Ref.
0309-1319-000):
•
Remove all air from the bulb.