S/5 Aespire
3-10
1009-0307-000
Set an audible alarm for circuit leaks
The patient circuit leak alarm comes on if less than half of the inspired volume
returns through the expiratory flow sensor during mechanical ventilation.
To prevent nuisance alarms from a known circuit leak (e.g. an un-cuffed
endotracheal tube), you can set the audio to off.
Normal volume and apnea monitoring does not change.
N
N
N
Nooo
otttteeee:::: This alarm is the first stage in detecting a circuit disconnect. You cannot
turn off the audible circuit leak alarm unless: volume alarms are on, and the
low
◊
E
limit is set to a value other than off.
Step 1
Push the Menu key to see
the main menu.
Step 2
Turn the knob to select
Alarm Settings.
Step 3
Push the knob to show
the next screen.