Set Cardiac Bypass
Cardiac Bypass (No / In Progress) turns off volume and apnea alarms when
these are not appropriate (e.g. during heart / lung bypass).
Set Cardiac Bypass to In Progress to prevent volume and apnea alarms when
the patient is on cardio-pulmonary bypass. When Cardiac Bypass In Progress
is selected, the display shows:
• Cardiac Bypass
• Apnea Alarm off
N
N
N
Nooo
otttteeee:::: Mechanical ventilation must be off. When mechanical ventilation is
turned back on, Cardiac Bypass returns to the No setting and monitoring is
available.
Step 1
Set the Bag/Vent switch
to Bag position:
R
This selects manual
ventilation (bag) and
stops the mechanical
ventilation (ventilator).
Step 2
Push the Menu key to see
the main menu.
Step 3
Turn the knob to select
the Alarm Settings.