S/5 Aespire
2-6
1009-0307-000
Figure 2-3 shows these controls on the front of the S/5 Aespire.
Item
Description
1
Auxiliary Common
Gas Outlet switch
(ACGO)
Sends fresh gas to the ACGO when the switch is
activated. The ACGO may be used to provide fresh
gas to an external manual breathing circuit.
Mechanical ventilation is not available when the
auxiliary outlet is selected and a medium priority
alarm will sound and a message “Aux Gas Outlet On”
is displayed. Pressure and volume monitoring are
not available when the ACGO is selected. Fresh gas
oxygen monitoring is available when the ACGO is
selected.
The ACGO should not be used to drive external
ventilators or for jet ventilation.
Breathing system selected:
Auxiliary outlet selected:
4 & 7
Flow sensor or flow
port adapter
Flow sensors provide volume measurements for
some monitoring functions. Flow port adapters have
no measurement capabilities.
Systems with volume monitoring but without volume
compensation have a flow port adapter in the front
position of the flow sensor module (inspiratory) and
a flow sensor in the rear position (expiratory).
Systems without volume compensation and without
volume monitoring have flow port adapters in both
positions of the flow sensor module. All other
systems have flow sensors in both positions of the
flow sensor module.