3 Operation and Tutorial
1009-0307-000
3-23
Connecting Active
AGSS without a flow
indicator
The active AGSS option without a flow indicator is for use only with the active
adjustable flow scavenging. The flow rate is limited to 30 L/min with this
option.
To use the optional active AGSS installed on the system which does not have
a flow indicator, connect it as follows.
1. Connect a disposal hose to the barbed connector on the needle valve on
the bottom of the AGSS below the ventilator engine (bellows).
•
The hose should be flexible and reinforced to help prevent kinking and
crushing.
•
Attach the other end of the hose to the hospital disposal system.
2. Attach the 3 liter visual
indicator bag to the 30 mm
auxiliary #1 port on the
bottom of the AGSS (below
the ventilator engine
(bellows).
3. Use the needle valve to adjust the flow rate. Use
the visual indicator bag when adjusting the flow
rate. The bag should be properly inflated to
indicate flow
.
4. Before you use the system, complete the
Preoperative Test procedure. Refer to section 4
Preoperative Tests of this manual.