4 Preoperative Tests
1009-0307-000
4-11
7. Test the Link 25 system with flow decreasing:
•
Set the N
2
O flow to 9.0 L/min.
•
Set the O
2
flow to 3 L/min or higher.
•
Slowly turn the O
2
flow control clockwise.
•
Make sure that the N
2
O flow decreases. The measured O
2
concentration must be
≥
21% through the full range.
8. Adjust the flow of all gases through their full range and make sure that the
flowtube floats move smoothly.
9. Disconnect the O
2
pipeline supply or close the O
2
cylinder valve.
10. Make sure that:
•
The low O
2
supply alarm occurs.
•
N
2
O, and O
2
flows stop. The O
2
flow stops last.
•
Air flow continues.
•
Gas supply alarms occur on the ventilator if the ventilator uses O
2
as
the drive gas.
11. Turn all the flow controls fully clockwise (minimum flow).
12. Reconnect the O
2
pipeline supplies or open the O
2
cylinder valve.
13. Set the system switch to Standby.
Vaporizer installation
w
w
w
w
WARNING
Use only the Selectatec series vaporizers Tec 4 or greater.
Do not use a vaporizer that lifts off the manifold when the lock
lever is in the locked position.
Do not use this anesthesia system if you can turn ON more than
one vaporizer at the same time.
Tec 6 Plus Vaporizers will not align correctly unless the power
cable goes through the channel on the bottom of the vaporizer. Do
not put the power cable on top of the manifold or between
vaporizers.
1. If the top of a vaporizer is not horizontal, remove the vaporizer and
reinstall it.