S/5 Aespire
3-26
1009-0307-000
Alarms
Alarms appear at the top of the screen. The highest priority alarm is shown in
Area 1, the next highest priority alarm in area 2. If all areas are used, the
lowest priority alarms cycle in area 4.
Alarm Tones
Alarm tones tell you about the alarm priority:
• High Priority: 2 bursts of 5 tones, every 10 seconds, (repeat)...
• Medium Priority: 3 tones every 25 seconds, (repeat)....
• Informational: single tone.
Alarm Silence
• The screen shows the time remaining in the silence period.
• If you push the alarm silence key while an alarm is silenced, countdown time
resets to 120 seconds:
• High priority alarms always cause an audible tone and must be silenced
individually.
Alarm Suspend
Pressing and holding the alarm silence key for one second, when no alarms
are active, suspends audio tones for medium and low priority alarms for 90
seconds.
Refer to Part 2 of this manual for more information about alarms and
messages.
B2
90
2
2
113
Area 1
00:01
Area 2
00:15
Area 3
00:30
Area 4
01:25
120