Datacolor CONDITIONER™ (MB2) 06/2017
page 76/126
11.4 Tolerance range settings
In this menu you can set the deviation between the actual value and setpoint which that shall cause a
tolerance range alarm.
This function only activates after the set-point has been reached once.
Path:
Main menu > Settings > Various
Submenu “Various”.
•
Select the field “Range alarm delay” and enter the time in minutes, after which the range alarm shall
be triggered. Setting range: 15 min to 120 min. Confirm entry with
Confirm
icon.
•
Select the field ”Temperature range” and enter the desired value for the temperature range. Setting
range: 2 °C to 10 °C. Confirm entry with
Confirm
icon.
•
Select the field “Humidity range” and enter the desired value for the humidity range. Setting range: 5%
r.H. to 20% r.H. Confirm entry with
Confirm
icon.
After completing the settings, press the
Confirm
icon to take over the entries and exit the menu,
or
press
the
Close
icon to exit the menu without taking over the entries.
If there are actual values outside the tolerance range the following information icons for the correspond-
ing parameter are displayed:
Icon
Signification
Information
“Temperature range”
The temperature value is outside the tolerance range
“Humidity range”
The humidity value is outside the tolerance range
If the condition persists, an alarm is triggered after the configured interval (“range alarm delay”). It is visu-
ally indicated in Normal display. If the alarm buzzer is activated (chap. 11.5) there is an audible alert. The
alarm is shown in the list of active alarms (chap. 11.3).
Содержание 9020-0129
Страница 124: ...Datacolor CONDITIONER MB2 06 2017 page 124 126 23 7 Dimensions size 115 ...
Страница 125: ...Datacolor CONDITIONER MB2 06 2017 page 125 126 24 EU Declaration of Conformity ...
Страница 126: ...Datacolor CONDITIONER MB2 06 2017 page 126 126 ...
Страница 127: ......