5.8
Performance when opening the door
When you open the door the fan starts running with minimum speed.
After 60 seconds from opening the door, heating, refrigeration, humidification, dehumidification and fan
turn off.
After closing the door, heating, refrigeration, humidification, dehumidification and fan turn on again.
6. Start up
6.1
Turning on the cabinet
After connecting the supply lines (chap. 4), turn on the cabinet by its main power switch (1). The lit pilot
lamp shows the cabinet is ready for operation.
When the main power switch is turned on and yet the controller display is dark, the display is in stand-by
mode. Press on the touchscreen to activate it.
•
Open the water-tap for freshwater supply. Alternatively, fill the freshwater can (option, chap. 4.5).
•
The humidifying and dehumidifying system must be activated (deactivated operation line “Humidity
off”, chap. 7.3 and setting “Control on”, chap. 6.3).
After the first turning on of the cabinet or after an interruption of the power supply the relative humidity will
increase after a delay of about 20 minutes, as the fresh water needs to be heated first. During this period,
the relative humidity can drop considerably.
The cabinet may release odors in the first few days after commissioning. This is not a quality defect. To
reduce odors quickly we recommend heating up the cabinet to its nominal temperature for one day and in
a well-ventilated location.
WARNING: If customer should use a cabinet running in non-supervised continuous opera-
tion, we strongly recommend in case of inclusion of irrecoverable specimen or samples to
split such specimen or samples and store them in at least two cabinets, if this is feasible.
6.2
Controller settings upon start up
The window „Language selection“ enables the
language selection
, in case that it’s activated in the
“Start-up” menu. Afterwards occurs a request of the
time zone
and the
temperature unit
.
The controller will function in the
operating mode
, which was active before the last shut-down. It controls
temperature and humidity in fixed value operating mode to the last entered values and in the program
mode to the set points achieved beforehand.