Datacolor CONDITIONER™ (MB2) 06/2017
page 31/126
To completely separate the cabinet from the power supply, you must disconnect the power
plug. Install the cabinet in a way that the power plug is easily accessible and can be easily
pulled in case of danger.
Remark when operating the cabinet with a power frequency of 60 Hz:
WARNING
High leakage current.
Electrical hazard.
Earth connection is essential before connecting supply. Check socket before inserting
plug.
When connected to a power supply 1N~ with a frequency of 60 Hz, a leakage current of more than 3.5
mAmp is possible. If grounding through the power cable is insufficient or missing, the leakage current
may flow through the user’s body. Correct installation of the professional grade power socket provided by
the user safely avoids this. Before connecting the chamber to the socket, please check its grounding
contact type plug for appropriate construction and if it is undamaged.
4.8
Connection of the optional voltage changer
The voltage changer enables the cabinet to operate at a power frequency of 115 Volt. It is packed sepa-
rately and supplied together with the cabinet.
The voltage changer is supplied with a fixed power connection cable with a NEMA 5-20P plug. It is pro-
tected against excess-current with an internal over-current release category B16A. The connection is
made by the customer.
CAUTION
Sliding or tilting of the voltage changer.
Damage to the voltage changer.
Risk of injury by lifting heavy loads.
Lift the voltage changer at both carrying handles from the pallet with two persons.
Do not install the voltage changer in the exhaust air flow at the rear of the cabinet.
For the installation of the voltage changer next to the cabinet, provide a wall distance the cabinet of ap-
prox. 0.4 m /
1.3 ft.
CAUTION
Danger of overheating.
Damage to the voltage changer.
∅
Do NOT install the voltage changer in unventilated recesses.
Ensure sufficient ventilation for dispersal of the heat.