Shad-o-Scan User Manual, Rev04
40
Shad-o-Scan 3001/4501 X-Ray Detector
Teledyne DALSA
The Rad-icon Detector Image Processing Control, as shown by CamExpert, groups
the image processing specific parameters. Defect Pixel Correction (DPC) and Flat
Field Correction (FFC) functions are supported by the detector. The calibration
reference maps required for DPC and FFC operations are loaded in the detector’s
non-volatile memory, for each
Readout
. A set of factory maps and a set of customer
maps are available. The factory maps are read-only while the customer maps can be
updated by the User.
The DPC and FFC functions can be used in combination with different
Trigger
and
Region of Interest
readout modes.
The function features are listed in the table below; the following paragraphs describe
these features in more detail.
Display Name
Feature
Acc Description
Factory
default
Min
Max
FFC Enable
FFCEnable
RW Turns On/Off Flat Field Correction
On
-
-
Off
Off
Off: Disabled
-
-
-
On
On
On: Enabled
-
-
-
FFC Processing
FFCActive
RO
Shows if FFC processing is enabled and
activated
-
-
-
Not Active
Not Active
FFC processing not active
-
-
-
Active
Active
FFC processing active
-
-
-
FFC Mode
FFCMode
RW Changes Mode for Flat Field Correction
Normal
-
-
Normal
Normal
Default Flat Field Correction Mode
-
-
-
New Dark
New Dark
Writes New Dark for Flat Field Correction
-
-
-
FFC New Dark
Weight
FFCNewDarkWe
ight
RW Select Weight for Dark reference update
10
0
100
Dark Map Save to
Flash
FFCNewDarkSa
veFlashReq
RW Save Dark Map to Flash
-
-
-
FFC NewDark
Status
FFCNewDarkSta
tus
RO
-
-
-
Map Select
FFCMapSelect
RW Select automatic or manual FFC map
selection
Auto
-
-
Auto
Auto
Automatic set FFC maps
-
-
-
Manual
Manual
Manual set FFC maps
-
-
-
Factory/Customer
Map Select
FFCFileSetSelect
CustomerMode
RW Indicates the mode of Customer or
Factory map
Auto
-
-