Shad-o-Scan User Manual, Rev04
37
Shad-o-Scan 3001/4501 X-Ray Detector
Teledyne DALSA
External Triggered Mode (Slave)
In this mode the detector will take a single image at the moment a trigger signal is
applied to the detector.
The falling edge of the trigger input defines the start of the readout time (B).
When using pulsed X-rays, the X-ray exposure including X-ray rise- and fall-times
have to be wi
thin the “Expos
ure Time
” window
(C).
The trigger output indicates the Exposure Time window when no read-out of the
sensor occurs.
Row #
Pixel Array
Time
ReadOut
Integration
Time
Frame (N)
Frame (N+1)
Trigger Input
Frame Blanking
interval
Sensor Data Output
C
External Trigger Mode
A
Operation mode
Binning 1x1
3001
Binning 1x1
4501
Binning 2x2
3001
Binning 2x2
4501
Binning 4x4
3001
Binning 4x4
4501
Max frame rate
1403 FPS
970 FPS
2703 FPS
1950 FPS
2703 FPS
1950 FPS
A: min integration
time
0.712 ms
1.014 ms
0.369 ms
0.513 ms
0.369 ms
0.513 ms
B: readout time
0.712 ms
1.014 ms
0.369 ms
0.513 ms
0.369 ms
0.513 ms
C: frame blanking
interval (A-B)
A-B
A-B
A-B
A-B
A-B
A-B
Figure timing diagram 'Trigger Mode'