Installation Precautions
Never connect or disconnect cables while power to the detector is on. Damage to the
detector electronics might occur if the cables are connected and/or disconnected while
the detectors power is on.
Service by Qualified Personnel Only
This equipment is to be installed by a qualified technician only. The system can only
be used in conjunction with a properly installed X-ray source with the appropriate
shielding and a properly configured computer workstation that meets the minimum
system requirements discussed later in this manual.
Environmental Specifications
Temperature
-
Operating
-
Transport & Storage
+10..+40
o
C
-10..+55
o
C
Humidity (non-condensing)
-
Operating
-
Transport & Storage
10%...80%
10%...80%
Shock
25G (6ms)
Vibration
1G (10-150Hz)
IP Class
IP69K
Note: For transport and storage, the standard product packaging is assumed to be in
place which includes a sealed anti-static shield bag.
Regulatory Compliance
The Shad-o-Scan has been designed and manufactured in accordance with the
applicable clauses of the international standards and legislations.
•
EN 61000-6-3: Electromagnetic compatibility (EMC)
–
Generic standards.
Emission standard for residential, commercial and light-industrial
environments
•
IEC/EN 61000-6-2: Electromagnetic compatibility (EMC)
–
Generic standards.
Immunity standard for industrial environments
•
EN 50581/EN IEC 63000: Technical documentations for the assessment of
electrical and electronic products with respect to the restriction of hazardous
substances
Detector Maintenance
Before performing any of the following operations, make sure that the power to the
detector is switched OFF.
Protecting the Detector Front Cover Carbon strip
The carbon strip in the front cover of the detector should be treated with care, as
scratches or debris in this area may produce artifacts in the X-ray image.