User Manual, Rev 01
7
Teledyne DALSA
Rad-icon 2329 GigE Detector
X-ray exposure should be limited only to the active image area (see chapter 5
mechanical drawing) in order not to cause damage to the electronic components of
the device.
Static precautions
Observe proper ESD/static control procedures when handling system components.
The use of properly grounded wrist strap is highly recommended.
Installation Precautions
Never connect or disconnect cables while power to the detector is on. Damage to the
detector electronics might occur if the cables are connected and/or disconnected
while the detectors power is on.
Enclosed Use
This equipment is designed for use in enclosed systems, with the enclosure
sufficiently shielding the equipment against the influence of visible light. If used
without enclosure and in bright environments, ambient light entering the detector
may impact image quality or the effectiveness of calibrations. It is the customer’s
responsibility to take measures in the application to avoid ambient light entering the
detector.
Service by Qualified Personnel Only
This equipment is to be installed by a qualified technician only. The system can only
be used in conjunction with a properly installed X-ray source with the appropriate
shielding and a properly configured computer workstation that meets the minimum
system requirements discussed later in this manual.
Environmental Specifications
Temperature
-
Operating
-
Transport & Storage
+10..+40
o
C
0..+50
o
C
Humidity (non-condensing)
-
Operating
-
Transport & Storage
20%..80%
20%..80%
Shock
25G (6ms)
Vibration
1G (10-150Hz)
Note: For transport and storage, the standard product packaging is assumed to be in
place which includes a sealed anti-static shield bag.