User Manual, Rev 01
31
Teledyne DALSA
Rad-icon 2329 GigE Detector
Detector Control
The Rad-icon Detector controls, as shown by CamExpert, groups the detector specific
parameters.
The features which are underlined link to the paragraph which describes these
features in more detail.
Display Name
Feature
Acc Description
Settings Revision
SettingsRevision
RO Displays the current version of the power
settings for driving the Sensor.
Standby
Standby
RW Enables or Disables the Standby. A few seconds
needed after coming out of the standby. Please
consult Teledyne Dalsa for more details on
image behavior after standby.
Linearity Equalizer
LinearityEqualizer
RW Enables or Disables the linearization
ReadOutMode
RW Select the way the sensor is read out
FullFOV 1x1
Full active pixel area
Full Field-of-View image (14-bit/pixel)
FullFOV 2x2
Binned full active area
Full Field-of-View image, 2x2 pixel binning
(15-bit/pixel)
FullFOV 1x2
Vertical binned full
active area
Full Field-of-View image, 2x vertical pixel binning (14-
bit/pixel)
Trigger Mode
RW Controls by what source initiate and how the
sensor is read out
Free Running
FreeRunning
Detector uses internal timing
Triggered
ExtTrigger
Read out of sensor occurs on falling edge of trigger
input
Snapshot
Snapshot
Sensor reset occurs on rising edge and sensor read out
occurs on falling-edge of trigger input
Snapshot Extended
Exposure
Timed Snapshot
Sensor reset occurs on rising edge of trigger and sensor
read out occurs after extended exposure counter reaches
its counter value
Triggered Sequence
Sequence of images
Sensor reset occurs on rising edge of trigger and then
records a number of exposed frames and offset frames
(trigger out is indicating the exposed frames).
Software Trigger
Trigger activated by
command
Sensor will start read out/reset depending on
synchronization mode
Trigger out
Control of trigger
output
When disabled trigger out is suppressed
Extended Exposure
ExtendedExposure
RW Time (milliseconds) in between two successive
read outs in Free Running Trigger Mode.
Extended exposure setting should not be
changed when the detector is making a
transaction. 377ms, 188ms and 63ms should be
used as minimum exposure time for FullFOV
1x1, FullFOV 1x2, FullFOV 2x2 modes
subsequently when the summing mode is 1x1
Please contact us for more information.
Offset
Offset
RW Sensor dark ADC offset level; based on full scale
of sensor
0%
Offset0pct
0
0.75%
Offset0pct75
0,75% of 16384 ADC full scale.
1.5%
Offset1pct5
1.5% of 16384 ADC full scale.
2.35%
Offset2pct35
2.35% of 16384 ADC full scale.