User Manual, Rev 01
35
Teledyne DALSA
Rad-icon 2329 GigE Detector
Row
#
Time
Reset
Integration
Time
Frame
(
N
)
Frame
(
N
+
1
)
Trigger Input
Trigger Output
Image array readout
A
C
Pixel Array
1
2
Readout
D
B
3
4
E
Full FOV 1x1
Full FOV 2x2
Full FOV 1x2
A
Trigger input active
B
Detector read out
125ms
62.5ms
62.5ms
C
X-ray active
Trigger input - D
Trigger input - D
Trigger input - D
D
Dummy read (reset)
= B
= B
= B
E
Minimum period time
500ms
125ms
250ms
Timing diagram in Snapshot mode
Snapshot Extended Mode.
In snapshot mode an image is taken, including a reset before integration.
Before X-ray is active the detector is read out to clear the pixels.
The detector is in slave mode. At rising edge of the trigger input signal a dummy read
(reset) is initiated. When the dummy read is finished, trigger out will be active.
The integration time is programmed by the Extended Exposure setting.
When the integration-timer is finished, trigger out will become inactive and the
detector read out starts. See figure below.