10.Troubleshooting
10.1 Fault Alarm Type
The processor’s fault type is divided into 3 levels
Table 6: Fault Alarm Type Table
Serial
No.
Type
Fault alarm description
Alarm mode
1
Danger
Imminent danger, if not avoided, it will cause
death or serious injury.
Monitor d
alarm ind
voice reminder
2
Warning
Potential risks, if not avoided, it will cause
serious injury or severe damage to the machine.
Monitor d
alarm ind
voice reminder
3
Note
Matters requiring attention or help information.
Monitor display
10.2 Troubleshooting
When HP300 Plus fails, take the following steps for processing:
Read error messages.
Operate according to the description of the message prompt.
Check the running log for fault message history. Perform troubleshooting according
to the fault message history.
Check the system’s quality control records for any parameter anomalies. Feedback
the parameter anomalies to the Dakewe Technical Service Department.
10.2.1 Power Failure
If the machine is unable to be turned on, perform a power check as follows:
Check whether the power plug is inserted into the socket and whether the
socket is powered on;
Make sure that the current leakage protection switch on the back of the
instrument is closed.
If you still cannot solve the problem through the above steps, contact Dakewe Technical
Services.