Dakewe (Shenzhen) Medical Equipment Co., Ltd
89
replacement. The user cannot start the cleaning program without replacing the cleaning
reagent. In addition, users can set the threshold of cleaning reagents in the reagent
quality control function according to the actual situation of the department. If the
number of uses of the cleaning reagent has reached the use threshold, the reagent
expired mark will be displayed on the reagent bottle, and the user needs to replace the
cleaning reagent.
Table 4: Usage Thresholds for Rinse Reagents
Serial No.
Reagent for rinse
Threshold value
2
Alcohol for rinse
12
3
Xylene for rinse
12
After reaching the usage threshold of rinse reagents, continuous use of the machine for
rinse will result in poor rinse effect and may cause damage and malfunction to the
instrument.