Dakewe (Shenzhen) Medical Equipment Co., Ltd
9.3 Regular Maintenance
Table 5: Maintenance Table
Serial
No.
Maintenance
Day
Week Month
Feb.
1
Use a piece of lint-free cloth to dip recommended
detergent to clean the outer surface of the
instrument.
√
2
After each rinse program of the dehydration
cylinder, dry the dehydration cylinder and cylinder
lid.
√
3
Check and clean the liquid level sensor.
√
4
Check the paraffin liquid level. Refill when
necessary.
√
5
Check the filling liquid level in the reagent bottle.
√
6
Check whether the reagent bottle has been fully
inserted into the connector.
√
7
Clean the waste paraffin collection tray.
√
8
Remove the paraffin residuals on the internal
surface of the paraffin cylinder. Check the paraffin
cylinder filter.
√
9
Check and empty the waste liquid collection bottle.
Clean the air intake.
√
10
Check the dehydration cylinder filter for any tissue
or paraffin residue.
√
11
Check and clean the seal ring of the dehydration
cylinder lid.
√
12
Clean the monitor.
√
13
Clean the interior of the reagent bottle. Lubricate
the seal ring of the reagent bottle. Check for
damage. Make sure that the protective cap and seal
ring of the reagent bottle are covered and tightly
closed.
√
14
Check the seal ring of the paraffin case door. Clean
the paraffin case.
√
15
Replace the active carbon filter.
√