8.Start Program
There are several conditions that must be met to start the program. In order to be able to start
the program, the following steps are very important.
Before turning on the instrument, visually inspect the instrument and accessories for any
damage that occurs while in transit or moving the instrument. If damage is found, do not
operate the instrument and contact your local service agency.
After transporting or moving the instrument, perform a pilot run (preloading program
or user-created program) before dehydrating the patient tissue samples to ensure the
suitability of the instrument and program used. After successful implementation of all
the above tests, the patient tissue samples used for diagnosis can be dehydrated. If any of
these steps fails, contact your local service agency.
Check the reagent liquid level and paraffin height before starting the program. Be sure
to keep the line of sight at the same height as the reagent liquid level / paraffin to ensure
that the reagent / paraffin volume does not exceed the maximum liquid level, nor does it
fall below the minimum liquid level. Insufficient reagents / paraffin may cause the
sample to dry out because it is not sufficiently submerged.
Check whether the reagent bottle and paraffin cylinder are inserted correctly.
Make sure that the reagent bottle site in the display interface matches the reagent bottle
in the reagent compartment correctly.
Check whether the dehydration cylinder strainer is properly installed at the bottom of
the dehydration cylinder and that the dehydration cylinder lid is sealed before the
program starts.
1. Confirm the prompt message in the main interface: The device functions properly and
paraffin has melted.
2. Click the “Open” button in the system interface. The electronic lock automatically opens
the dehydration cylinder lid (See Section 4.4.2).
3. Put the sample basket into the dehydration cylinder.
Use a tissue processing basket during dehydration. Do not place the sample directly into
the dehydration cylinder.
Handle the sample carefully and place it in the cassette so that it does not fall into the
holes on the dehydration cylinder strainer.
4. After confirming the number of cassettes and tissue processing baskets, close the
dehydration cylinder lid. Click the “Close” button in the system’s main interface again.