Dakewe (Shenzhen) Medical Equipment Co., Ltd
35
4.3.14 Cleaning xylene pre-heating
When the ambient temperature is low, the compatibility between paraffin and
xylene is low in the running state of the equipment, and there is a potential
risk of pipe blockage when the xylene is not replaced in time. Therefore, it is
particularly important to preheat and clean xylene in advance.
1.Temperature sensor: real-time detection of cleaning xylene reagent
temperature
2. Heating plate: provide heat source, and overheat protection
3. Aluminum base: perfectly fits to the bottom of reagent bottle for maximum
thermal conductivity
4. Heat insulation: prevent overheating loss of other reagents
4.4 Alarm Function
In the instrument, there will be some situations that require the user's attention or
decision. The simplest case is some confirmation operations to continue the dehydration
program. In addition, in the continuous monitoring of the hardware, there will be errors
that must be eliminated as soon as possible in order to successfully complete the
program operation. Accordingly, all messages will be classified according to their
severity.
HP300 Plus has two different types of alarm functions:
Native alarm
This machine audible alarm is sent from the inside of the machine and is used for all alarm
information during the operation of the machine.
Remote alarm
This alarm type is located outside the instrument. If the device is connected to a network
(Figure 22-1), an alert message will be sent over the network to the responsible personnel.
The remote alarm can be received by electronic devices such as computers, IPADs or mobile