6.2.1.4 Function settings
Turn on the paraffin melting function
(Figure 55-1)
: refer to "Paraffin melting function" in "Boot
Setup Preference 2" in Section 6 of Chapter 4.
Turn on the cleaning and baking function
(Figure 55-2)
: See "Cleaning and Baking Function" in
"Boot Setup Preferences 2" in Section 6 of Chapter 4.
Enable the reagent replenishment function
(Figure 55-3)
: Touch the check box to enable the
reagent replenishment function. When the same reagent repeated used during the operation of the
program and the current reagent is insufficient, the liquid replenishment will be performed. For
example, both S7 and S8 are 100% ethanol, if the reagent level in the reagent tank at S7 site is less
than the level required by the retort, the reagent will be extracted from the reagent tank at S8 site
and added to the retort , after soaking, all reagents will be discharged to the reagent tank at S7 site.
Enable the reagent automatic rotation function
(Figure 55-4)
: The reagent automatic rotation
function is enabled by default. Tap to enable the reagent automatic rotation function to enable the
following three function modes. 1. Drain out old liquid (Step 1: pumping liquid from the reagent
bottle to the retort, Step 2: drain liquid from the retort to the waste liquid tank, Step 3: clean the
draining pipe (blow air into the draining pipe for 30S) 2. Refill new liquid (Step 4: refill liquid
from the new liquid bottle to the retort, Step 5: drain liquid from the retort to the reagent bottle,
Step 6: clean the filling pipe (blow air into the filling pipe for 30s) 3. One-touch reagent rotation
(drain old liquid first, then refill new liquid)
Enable the cleaning xylene preheating function
(Figure 55-5)
: After the cleaning xylene preheating
function is enabled, click Save to set the preheating temperature (40-55
℃
). The heating sign and
real-time preheating temperature can be displayed at site C1 on the main panel.
Enable concentration detection function
(Figure 55-6)
: The concentration detection function is
enabled by default. After this function is enabled, HP300 Plus will automatically detect and update
the concentration of ethanol reagent in the first step of dehydration process, automatic liquid
change and fresh liquid extraction. You can view the real-time ethanol concentration and the preset
concentration threshold on the main panel. For details about how to set the concentration threshold,
see Section 3 of Chapter 6. Set threshold in reagent quality control.