Table II: Electromagnetic immunity
Electromagnetic immunity
Immunity test items
Basic standards
Test values
Performance
criteria
Electrostatic discharge
(ESD)
IEC61000-4-2
Contact discharge: ±2kV, ±4kV
Air discharge: ±2kV, ±4kV, ±8kV
B
Radio frequency
electromagnetic field
IEC61000-4-3
3V/m (80 MH
z
-1GH
z
)
3V/
m
(1.4GH
z
-2GH
z
)
1V/m(2.0GH
z
-2.7GH
z
)
A
Pulse cluster
IEC61000-4-4
Power cord: ±1kV(5/50ns,5kHz)
B
Surge
IEC61000-4-5
Line to ground: ±2kV
Line to line: ±1kV
B
Radio frequency
transmission
IEC61000-4-6
Power cord: 3V, 150kHz- 80MHz,
80%AM
A
Power frequency magnetic
field
IEC61000-4-8
3A/m, 50/60Hz
A
Voltage sag and interruption
IEC61000-4-11
1 cycle, 0%
5/6 cycle, 40%;
25/30 cycle, 70%;
250/300 cycle, 5%
B
B
C
C
Performance criteria:
A. In the test, the performance is normal within the specified limit.
B. In the test, the function or performance is degraded or lost temporarily, but can restore automatically.
C. In the test, the function or performance is degraded or lost temporarily, and cannot restore until reference or
system operation by the operator