35
(2) Select a sensor to temporarily lower the setting value by scrolling the screen with the upward and
downward arrow keys.
Select the sensor and press the [ENTER] key to define the selection.
(3) Set the HIGH ALARM level for the selected sensor.
The value currently detected is displayed (5 is displayed as an example here), and the value of about 10
above the currently displayed value (15 as an example here) is automatically set as the HIGH ALARM
level.
Increase or decrease the automatically set value as required with the upward and downward arrow keys and
press the [ENTER] key to define the final value.
(4) The currently detected value and set value are displayed. Select the
「
1. CHECK
」
and press the
[ENTER] key to start the test.
(5) The
「
ALARM OPERATION OK
」
message is displayed during the test process, showing the detector is
in the test process.
Press the [ENTER] key to suspend the test process if the process must be stopped.
(6) Check the
「
PRE ALARM
」
and
「
HIGH ALARM
」
messages are displayed if the mist density exceeds
the set level at the specified sensor.
Press the [ENTER] key to return to the normal operating screen.
Original content from www.daihatsu.cc
Issued to HORIZON TANKERS LIMITED S.A.