77
Resolution Test Phantom
Carpal support position
When it´s ready press OK, the software will start a 150-second countdown.
During the countdown take an exposure. The software will show a picture
below.
Maximum Contrast Resolution
Verify if the lines of the contrast element are visible.
14.4. QA DIAGNOSIS FOR TOMOGRAPHY
14.4.1. PREPARING X-RAY UNIT
This procedure will produce tomographic images. It will be necessary take
an exposure 8x8 HD mode and 8x12 HD mode.
The software will indicate which phantom should be prepared.
a) Small FOV
Place the Small FOV QA phantom on chin rest support and level it.
Fill water the density phantom