14.3. QA DIAGNOSIS FOR CEPHALOMETRIC
This procedure will produce cephalometric images.
14.3.1. BEAM POSITION
The software will indicate that the beam position will be checked.
Remove the chin rest support, open the head holder. Rotate the cephalostat
to lock at PA/AP position and open the ear rods.
When it´s ready press OK, the software will start a 9-second countdown.
During the countdown take an exposure.
The following screen will be show.
14.3.2. CALIBRATION
The software will indicate that the calibration will be checked.
Rotate the cephalostat and lock at Lateral Cephalometric position.
When it´s ready press OK, the software will start a 150-second countdown.
During the countdown take an exposure. The software will show a picture.
Both ear holders have metal that appear in a clear way on the x-ray. The
validation consists of verifying if the sphere is inside the circle as shown in the
following picture: