9
3.5. ADDITIONAL PROCEDURES PRIOR TO EQUIPMENT USE
Even prior to its first use, the equipment must be cleaned and
disinfected; the same additional procedures must be followed for reuse, as
described in Chapter
3.6. WARNINGS AND/OR CAUTION TO BE ADOPTED
3.6.1. WARNINGS AND/OR CAUTION DURING EQUIPMENT
INSTALLATION
The equipment must be installed only by service technicians authorized
by the manufacturer.
Place the equipment on a site where it will not be in contact with
moisture or water.
Install the unit on a site where it will not be damaged by pressure,
temperature, moisture, direct sunlight, dust or salts.
The equipment must not be submitted to inclination, excessive vibration
or shock (including during transportation and handling).
This equipment has not been designed for use in facilities where
vapors, flammable anesthetic mixtures in contact with air, oxygen or
nitrous oxide can be detected.
Check the equipment’s voltage when performing electric installation.
Failure to do so may damage the equipment.
The equipment must be properly grounded.
Failure to do so may
result in “Safety Hazard”.
Depending on local regulation the X-Ray emission control may require
installation outside the facility where the equipment is placed, and the
operator may need visual contact with the patient through a window
with radiological glass or similar, since the operator must not lose visual
contact with the patient.
Mobile and portable RF communication equipment can affect the Eagle
3D CT Scanner.
This equipment must be solely used by health care professionals as it
may cause radio interference or interrupt the operation of nearby
equipment. Mitigatory measures, such as equipment re-orientation or
replacement and the facility’s screening, may be necessary.
3.6.2. WARNING AND/OR CAUTION DURING EQUIPMENT USE
The equipment must be operated only by qualified and trained
professionals (dentists, radiology technicians, hygienists or
engineers).
Always observe the display messages, the equipment as a whole
and the patient in order to detect any arising problems early.
In case occasional maintenance is required, use only services
provided by Authorized Service Technicians.