76
Calibration Check for Cephalometric
14.3.3. MAXIMUM CONTRAST RESOLUTION
The software will indicate that the maximum contrast resolution will be
checked.
Remove the chin rest support, open the head holder. Rotate the cephalostat
to lock at PA/AP position and open the ear rods.
Place the Resolution Test Phantom at carpal support and place it at the
cephalostat as shown below.