Step
Action
2
If the instrument is not powering down, disconnect the power cord from the
instrument.
Note:
The emergency shutdown only shuts down the instrument. Any computer
that is connected to the instrument will still be on. The connected computer
will enter sleep mode after a set period of time.
Power failure
The following table describes the consequences of a power failure.
Power failure to...
will result in...
ImageQuant 800
instrument
•
The run is interrupted immediately.
•
Any image capture in process will be lost. Only saved data
will be available on reboot.
•
The
Power indicator switches off.
Computer
•
The
Power indicator on the instrument control panel
displays blue.
•
The run is interrupted immediately.
•
If a run was in process, no image data will be saved.
Restart after emergency shutdown or
power failure
To restart the system after an emergency shutdown, follow the startup procedure in
Section 5.1 Start the Amersham ImageQuant 800, on page 71
2 Safety Instructions
2.3 Emergency procedures
Amersham ImageQuant 800 Operating Instructions 29372604 AC
25