CAUTION
Changing the filters must be done using the Amersham
ImageQuant 800 control software procedure. Do not manually
rotate the filter turret.
CAUTION
Do not use the instrument in a room with a temperature above
28°C. Higher temperatures do not allow the CCD to cool down
properly (to -25°C).
CAUTION
Do not insert or spill any substances into the ventilation hole on the
side of the instrument.
CAUTION
Reagents used to prepare the sample should be used in accord-
ance with the manufacturer's instructions.
CAUTION
Use the handle to open or close the side door.
CAUTION
Handle the sample trays and NP lens with care to avoid dropping
them.
CAUTION
Do not place heavy objects on top of the instrument.
CAUTION
Do not touch the light source in the instrument. The light source
may be hot.
2 Safety Instructions
2.1 Safety precautions
16