Parameter
Requirement
Transportation/storage
conditions
Temperature: -20°C to 70°C
Humidity: 5% to 95% RH (no dew condensation)
Operation site
Indoor use
Maximum operating
altitude
≤ 2000 m
Noise
< 60 dB(A) or lower
A-weighted sound rpessure level LAeq, t ≤ 60 dB, fast (t
= 0.25 s 1 M from the instrument)
Protection provided by
enclosure
IP21
Overvoltage category
Transient overvoltage category II
Rated pollution applied
Pollution degree 2
Note:
1. When the above conditions cannot be satisfied, take appropriate
actions.
2. Take into consideration the workflow and ancillary facilities when plan-
ning the installation.
3. Any required construction and electricity/air conditioning work must be
completed prior to installation.
4. It is not desirable to have a heat source near the Amersham
ImageQuant 800 air intakes, even if the other environmental require-
ments have been met.
5. Do not install the equipment near a window. Avoid direct sunlight. Make
sure that blinds are attached to nearby windows.
6. Do not place objects near the power outlet to allow easy access to the
power cord for disconnection in case of emergency.