Operation / Technical Manual
NORM-O-TEMP
®
, Model 111W
Page 63 of 82
Section 7.
Troubleshooting Guide
Observation
Possible Problem
Action To Be Taken
A. The power switch is
pressed to the ON position
but it does not light; control
panel is blank
Unit is unplugged.
Power switch is defective
Fuse is open
No line voltage
Check that the power cord is
plugged into a properly
grounded hospital grade
outlet.
Replace the power switch.
See Section 6-13.
Check the circuit breaker for
the outlet.
Check for possible open in
electrical system.
B. Each time the power
switch is pressed to the
ON position, it snaps
back to the OFF position
The fuses are defective.
Check amperage at switch.
If the current measurement
is less than the specified
amperage draw check for
electrical short in unit.
If the current measurement
is more than the specified
amperage draw check the
heater and pump amperage.
CAUTION
This device is still energized when the power switch is in the off position. To
completely disconnect the device from the power source, remove the power cord plug
from the power source.
Содержание Norm-O-Temp 111W
Страница 15: ...Operation Technical Manual NORM O TEMP Model 111W Page 15 of 82 Figure 1 NORM O TEMP unit Front View...
Страница 17: ...Operation Technical Manual NORM O TEMP Model 111W Page 17 of 82 Figure 2 NORM O TEMP unit Left Side View...
Страница 19: ...Operation Technical Manual NORM O TEMP Model 111W Page 19 of 82 Figure 3 NORM O TEMP unit Rear View...
Страница 21: ...Operation Technical Manual NORM O TEMP Model 111W Page 21 of 82 Figure 4 NORM O TEMP unit Right Side View...
Страница 23: ...Operation Technical Manual NORM O TEMP Model 111W Page 23 of 82 Figure 5 NORM O TEMP unit Model 111W Top View...
Страница 76: ...Operation Technical Manual NORM O TEMP Model 111W Page 76 of 82 Figure 14 NORM O TEMP Electrical Wiring Diagram...
Страница 77: ...Operation Technical Manual NORM O TEMP Model 111W Page 77 of 82 Figure 15 NORM O TEMP Water Circulation Diagram...