WARNING
A physician's order is required for the use of the device and setting the temperature of the
blanket/pad. At least every 20 minutes, or as directed by physician, check patient's
temperature and skin condition of areas in contact with blanket/pad; also, check
blanket/pad water temperature. Pediatric patients, temperature-sensitive patients with
vascular disease, surgical patients, diabetic and Reynaud’s disease patients should be
checked more frequently.
If patient’s temperature does not reach desired set point or
differs drastically from recommended set point, notify physician.
Notify the physician
promptly of any change in patient status in order to avoid serious injury or death.
1.
Pediatrics – Infant’s and children’s body temperatures are often more
responsive to surface heating and cooling than adults. Due to their size, the
effect of heating or cooling a child is likely more pronounced because of their
higher ratio of skin contact area to body mass.
2.
Temperature Sensitive Patients – Patients with impaired peripheral blood
circulation and patients who are incapacitated may be more sensitive to
temperature changes than patients with normal circulation.
3.
Surgical Patients – Patients with poor circulation due to inadequate heart
function, loss of blood, or impaired peripheral blood circulation may be more
sensitive to temperature changes.
The method of temperature control provided by all hyperthermia units presents the danger
of heating body tissues, particularly the skin, to a point where they are injured. Depending on
the extent and severity of a burn, very serious and even fatal complications may arise
. Avoid
placing additional heating or cooling sources between the patient and blanket/pad.
Depending on the extent and severity of a burn,
very serious and even fatal complications
may arise.
Do not use the NORM-O-TEMP
®
System distal to arterial cross clamping.
Thermal injury
may result.
Prevent excessive and/or prolonged tissue pressure and shearing forces, especially over
bony prominences,
to prevent skin damage that may result.
Do not place additional heat sources between the patient and blanket/pad.
Skin damage
may result.
The area between the patient and the blanket/pad should be kept dry to avoid injury
to patient.
Prep solutions have been reported to injure the skin when allowed to remain
between patients and a water-circulating heating blanket/pad during prolonged procedures.
If using a vinyl blanket/pad, place a dry absorbent sheet between the patient and the
blanket/pad to prevent moisture from accumulating.
Proper sanitation procedures must be practiced and hygienic safety must be
maintained, to prevent contamination.
Contamination can affect patient’s health, i.e. skin
irritation/rash may result.
Do not use the NORM-O-TEMP
®
system in the presence of flammable anesthetics.
Risk of
explosion can result.
Power interruption will cause the NORM-O-TEMP
®
to revert to Preset Set point Temperature
resulting in possible inaccurate therapy to the patient. Follow instructions for First Time Set-
Up/System Test Routine to resume operation.
Failure to resume therapy could result in
serious injury or death.
Содержание Norm-O-Temp 111W
Страница 15: ...Operation Technical Manual NORM O TEMP Model 111W Page 15 of 82 Figure 1 NORM O TEMP unit Front View...
Страница 17: ...Operation Technical Manual NORM O TEMP Model 111W Page 17 of 82 Figure 2 NORM O TEMP unit Left Side View...
Страница 19: ...Operation Technical Manual NORM O TEMP Model 111W Page 19 of 82 Figure 3 NORM O TEMP unit Rear View...
Страница 21: ...Operation Technical Manual NORM O TEMP Model 111W Page 21 of 82 Figure 4 NORM O TEMP unit Right Side View...
Страница 23: ...Operation Technical Manual NORM O TEMP Model 111W Page 23 of 82 Figure 5 NORM O TEMP unit Model 111W Top View...
Страница 76: ...Operation Technical Manual NORM O TEMP Model 111W Page 76 of 82 Figure 14 NORM O TEMP Electrical Wiring Diagram...
Страница 77: ...Operation Technical Manual NORM O TEMP Model 111W Page 77 of 82 Figure 15 NORM O TEMP Water Circulation Diagram...