Operation / Technical Manual
NORM-O-TEMP
®
, Model 111W
Page 62 of 82
6-17.1.
Ground Integrity
Using a Digital Multimeter, measure the resistance between an exposed, unpainted
part of the metal case of the NORM-O-TEMP
®
and the ground conductor of the
power plug. Take measurements between ground pin and the Hansen fittings, also
between the ground pin and the handle. Verify that the case to ground pin
resistance is less than 0.2 ohms.
6-17.2.
Leakage Current
Using an Electrical Safety Analyzer, measure electrical leakage current of the under
the following conditions: Power ON & OFF, polarity normal and reverse, unit
grounded and ungrounded. When measuring power ON, polarity normal and
reverse, measure when the NORM-O-TEMP
®
unit is heating and circulating water
(both the heater and pump are ON). When measuring power OFF, normal and
reverse polarity, measure while the NORM-O-TEMP
®
unit is going through self-
calibration, when it is first turned on (both the heater and pump are OFF). The
leakage current should be less than 500 microamps. If a NORM-O-TEMP
®
unit has
leakage current that exceeds the appropriate limit or has a significant increase in
leakage current, the cause should be investigated.
Excessive leakage current is most commonly caused by a defective heater, but other
components can also fail in such a way as to increase leakage current. To find the
source of excessive leakage current, suspect components should be sequentially
disconnected until the source of the high leakage current is isolated.
Содержание Norm-O-Temp 111W
Страница 15: ...Operation Technical Manual NORM O TEMP Model 111W Page 15 of 82 Figure 1 NORM O TEMP unit Front View...
Страница 17: ...Operation Technical Manual NORM O TEMP Model 111W Page 17 of 82 Figure 2 NORM O TEMP unit Left Side View...
Страница 19: ...Operation Technical Manual NORM O TEMP Model 111W Page 19 of 82 Figure 3 NORM O TEMP unit Rear View...
Страница 21: ...Operation Technical Manual NORM O TEMP Model 111W Page 21 of 82 Figure 4 NORM O TEMP unit Right Side View...
Страница 23: ...Operation Technical Manual NORM O TEMP Model 111W Page 23 of 82 Figure 5 NORM O TEMP unit Model 111W Top View...
Страница 76: ...Operation Technical Manual NORM O TEMP Model 111W Page 76 of 82 Figure 14 NORM O TEMP Electrical Wiring Diagram...
Страница 77: ...Operation Technical Manual NORM O TEMP Model 111W Page 77 of 82 Figure 15 NORM O TEMP Water Circulation Diagram...