Operation/Technical Manual
NORM-O-TEMP
®
, Model 111W
Page 12 of 82
Draining Instructions:
1. Turn “OFF” power switch.
2. Connect drain hose to one of the outlet fittings and allow to gravity drain.
Alarm Conditions:
If an alarm sounds, check indicator lights or display for problem and take corrective
action.
Alarm may be silenced temporarily for five minutes by depressing the “Silence Alarm”
button.
The “Silence Alarm” button will not silence the alarm when “ERR” appears on the
display or “HIGH TEMP” indicator light comes on and does not flash. Power cord
should be removed from outlet and the unit sent to Biomedical Engineering for repair.
CAUTION
:
This device should be operated only by trained personnel, under the direction of a
physician.
Patients vary in degree of sensitivity to cold, heat, and pressure.
The patient’s temperature and skin conditions should be checked at least every 20
minutes, or as directed by a physician, while on a thermal blanket.
Содержание Norm-O-Temp 111W
Страница 15: ...Operation Technical Manual NORM O TEMP Model 111W Page 15 of 82 Figure 1 NORM O TEMP unit Front View...
Страница 17: ...Operation Technical Manual NORM O TEMP Model 111W Page 17 of 82 Figure 2 NORM O TEMP unit Left Side View...
Страница 19: ...Operation Technical Manual NORM O TEMP Model 111W Page 19 of 82 Figure 3 NORM O TEMP unit Rear View...
Страница 21: ...Operation Technical Manual NORM O TEMP Model 111W Page 21 of 82 Figure 4 NORM O TEMP unit Right Side View...
Страница 23: ...Operation Technical Manual NORM O TEMP Model 111W Page 23 of 82 Figure 5 NORM O TEMP unit Model 111W Top View...
Страница 76: ...Operation Technical Manual NORM O TEMP Model 111W Page 76 of 82 Figure 14 NORM O TEMP Electrical Wiring Diagram...
Страница 77: ...Operation Technical Manual NORM O TEMP Model 111W Page 77 of 82 Figure 15 NORM O TEMP Water Circulation Diagram...