Operation / Technical Manual
NORM-O-TEMP
®
, Model 111W
Page 41 of 82
D.
Fill the reservoir with warm distilled water.
E.
Turn the unit on and circulate per the Chemical Cleaning Circulation Chart
(above).
F.
Drain the unit as instructed in Step A.
G.
Rinse the unit three (3) times as described in Steps D-F. Be sure to omit the
household bleach (sodium hypochlorite).
H.
After the third rinse, drain the unit and fill the reservoir with distilled water and
circulate.
I.
Check the water with pH strips or other appropriate test method for detecting
bleach. If bleach is detected, repeat Steps F-H.
J.
If unit is to be
returned to service, once no bleach is detected, add the
appropriate amount of U.S.P. Grade Propylene Glycol to the water reservoir per
the following chart:
UNIT
U.S.P. Grade Propylene Glycol Per Unit
NORM-O-TEMP
®
, 111W
12 ounces (340cc)
K.
Continue to fill the water reservoir with distilled water.
L.
If unit is being placed in dry storage skip steps J and K, continue with procedure
at step M.
M.
When all fluid has been removed from the unit, disconnect the drain hose(s) and
wipe unit clean.
N.
Unit is now ready for storage.
O.
Document unit maintenance.
When you are ready to return unit to service, replenish the reservoir(s) with
distilled water and 340 cc propylene glycol. Perform all the checks as described in
the respective Manual.
Содержание Norm-O-Temp 111W
Страница 15: ...Operation Technical Manual NORM O TEMP Model 111W Page 15 of 82 Figure 1 NORM O TEMP unit Front View...
Страница 17: ...Operation Technical Manual NORM O TEMP Model 111W Page 17 of 82 Figure 2 NORM O TEMP unit Left Side View...
Страница 19: ...Operation Technical Manual NORM O TEMP Model 111W Page 19 of 82 Figure 3 NORM O TEMP unit Rear View...
Страница 21: ...Operation Technical Manual NORM O TEMP Model 111W Page 21 of 82 Figure 4 NORM O TEMP unit Right Side View...
Страница 23: ...Operation Technical Manual NORM O TEMP Model 111W Page 23 of 82 Figure 5 NORM O TEMP unit Model 111W Top View...
Страница 76: ...Operation Technical Manual NORM O TEMP Model 111W Page 76 of 82 Figure 14 NORM O TEMP Electrical Wiring Diagram...
Страница 77: ...Operation Technical Manual NORM O TEMP Model 111W Page 77 of 82 Figure 15 NORM O TEMP Water Circulation Diagram...