Operation / Technical Manual
NORM-O-TEMP
®
, Model 111W
Page 29 of 82
B.
Visually inspect the NORM-O-TEMP
®
unit to determine that there are no missing
parts, dents, or punctures.
C.
Examine the power cord for cuts or exposed wires and the power plug for bent
or missing prongs.
D.
Review Section 1-3 to identify the features of the NORM-O-TEMP
®
unit.
E.
Collect and arrange the following equipment and supplies:
1.
Hyper-hypothermia blanket/pad(s) Listed in Figure 8.
2.
Connecting hose with quick-disconnect Hansen fittings (if applicable).
F.
Mount the NORM-O-TEMP
®
unit on the proper stand.
1.
To mount the NORM-O-TEMP
®
on IV Pole Stand:
i.
Place the NORM-O-TEMP
®
on the equipment pan with the rear of the
unit facing the IV pole. Slide the NORM-O-TEMP
®
toward the IV pole
until the rubber bumper touches the pole
ii.
Fasten the NORM-O-TEMP
®
by turning the knurled knobs (on the
underside of the equipment pan) clockwise into the threaded holes on
the bottom of the NORM-O-TEMP
®
. Hand tighten knurled knobs until
snug.
iii.
Fasten the mounting bracket around the lower IV pole near the
bumper on the rear of the NORM-O-TEMP
®
with the two stainless
steel machine screws provided.
2.
To mount the NORM-O-TEMP
®
on the Low Profile Stand:
i.
Place the NORM-O-TEMP
®
unit on the equipment pan so that the
threaded holes on the bottom of the unit line up with the knurled
knobs (on the underside of the equipment pan).
ii.
Fasten the NORM-O-TEMP
®
by turning the knurled knobs clockwise
into the threaded holes on the bottom of the NORM-O-TEMP
®
. Hand
tighten knurled knobs until snug.
G.
Slide the lid of the water fill opening and add 340 cc of propylene glycol then
gradually pour approximately 6 quarts (5.7 liters) of distilled water into the
reservoir. DO NOT USE DE-IONIZED WATER OR TAP WATER. Stop pouring
when the water reaches the strainer visible at the bottom of the water fill
opening. DO NOT OVERFILL.
H.
Connect the blanket/pad(s) to the NORM-O-TEMP
®
unit by attaching the quick-
disconnect female Hansen socket of the connecting hose to a male outlet Hansen
plug (on the bottom row) of the unit. Attach the male quick-disconnect Hansen
plug of the connecting hose to a female return Hansen socket (on the top row) of
the unit. Each blanket/pad must be connected to one outlet coupling and one
return coupling of the connecting hose if necessary.
1.
To attach the Hansen fittings (quick-disconnect fittings):
i.
Grasp the quick-disconnect female Hansen socket of the connecting
hose
Содержание Norm-O-Temp 111W
Страница 15: ...Operation Technical Manual NORM O TEMP Model 111W Page 15 of 82 Figure 1 NORM O TEMP unit Front View...
Страница 17: ...Operation Technical Manual NORM O TEMP Model 111W Page 17 of 82 Figure 2 NORM O TEMP unit Left Side View...
Страница 19: ...Operation Technical Manual NORM O TEMP Model 111W Page 19 of 82 Figure 3 NORM O TEMP unit Rear View...
Страница 21: ...Operation Technical Manual NORM O TEMP Model 111W Page 21 of 82 Figure 4 NORM O TEMP unit Right Side View...
Страница 23: ...Operation Technical Manual NORM O TEMP Model 111W Page 23 of 82 Figure 5 NORM O TEMP unit Model 111W Top View...
Страница 76: ...Operation Technical Manual NORM O TEMP Model 111W Page 76 of 82 Figure 14 NORM O TEMP Electrical Wiring Diagram...