1. Do not move the pressure adjustment rod on the regulator when changing the CO
2
cylinder. It will affect the secondary pressure and will result in needing to reset the
valve. The inline tube might fall off if the pressure is too high and maintenance will be
required.
2. Check tube regularly to ensure it is safely connected. Change the tube if it is
damaged or aged.
3. Use the stainless steel hoop to tighten the connection point of the Polyurethane tube
and the filter to prevent the pipe from falling off and leaking gas.
4. Use more than 99.99% pure CO
2
. Low purity may cause blocking of the filter or
damage of the electromagnetic valve.
5
.
Connection components
A. Schematic diagram
B. Actual picture
4. Check the tube connections for possible leakage (the connections between tube and
regulator, tube and access port, and tube and filter).
Note:
5. Installation
17
18
Polyurethane Pipe
(
Inner dia.7mm,
outer dia. 10mm)
To the filter
Stainless steel hoop
To the CO
2
regulator
6. Structure
●
Down-swing-door
●
Up-swing-door
Ventilation
Ventilation
Adjustable feet
Adjustable feet
Handle
Handle
Touchscreen Control Interface
WIFI
Interface
USB
Data Out
UV lamp switch
Power switch
Remote wifi monitor interface
(
optional
)
Co2 Gas Inlet
I
O
I
O
USB port
Warnings
- Make sure the gas supply meets the specified requirements.
- Make sure all the pipes are properly connected.
- Make sure the gas pressure settings are correct.
- Wrong pipe connection or pressure settings may cause CO
2
leakage.
- High Co
2
concentration will do harm to your health and may cause
suffocation or death.
- Alternative methods should be taken if there is no sufficient ventilation in the
room to guarantee safety. These methods include gas monitoring and alarming
system.
- Maintaining correct gas pressures can prevent gas leakage.