e.
To stack the 2nd or 3rd unit, fix it with four screws at the designated locations below
on both sides.
16
5. Installation
5. Installation
Please make sure each instrument is placed level, stable and still. Do not exceed the
maximum load of 20kg and the maximum stacked shaker speed of 250 RPM on the
second and third units.
Note
a. When stacked up to 3 units, units will need a Two-way Connector and 2 T-connectors.
b. When stacked up to 2 units, units will need a Two-way Connector and a T-connector.
5.3 Drain connection
2. Using the polyurethane tube provided,
connect the regulator to a CO2 inlet HEPA
filter first, and then to the CO2 inlet located
on the right side of the shaker. (See Figure 1)
3. Set the CO2 secondary flow control valve to 0.08~ 0.1MPa
(11.6~14.5psi).
(
WARNING
: Excessive pressure may cause disconnection
of internal pipes inside the CO2 incubator which will result
in leakage of CO2 gas into the surrounding environment.
High CO2 concentration can cause harm to your
health and may cause asphyxia or death. Maintenance
is required if the inline tube falls off.)
5.4 CO2 Connection
1. Install a regulator with flow control valve onto the CO
2
Cylinder. Please use a
2
2
regulator with a primary preset pressure of 25MPa (250kg/cm , 3500l b/in ), and a
secondary preset pressure of 0.6MPa (2.0kg/cm, 30psiG).
Gas Inlet
PU Pipe
Filter
Protection Cover
Figure 1
Figure 2
Two-way
Connector
T-connector
15
Содержание AEOLUS IS-18A
Страница 1: ...AEOLUS Stackable Incubator Shaker...
Страница 43: ...AEOLUS Stackable Incubator Shaker...