d: Humidity Calibrate
Humidity can be calibrated by following the procedure below:
1. Set the instrument to the target
humidity % (start with 25%)
2. Let instrument run and wait for the
humidity % to stabilize
3. Measure
humidity % inside chamber
4. Input the difference between set
humidity % and actual humidity % with “ ” or “ ”
5. Repeat steps 1-4 for 50%, 70%, and 90%
0
.
0
0
.
0
0
.
0
0
.
0
1
.
0
3
.
0
5
.
0
10
.
0
c: CO2% Calibrate
CO
2
% Concentration can be calibrated by following the procedure below:
1. Set the instrument to the target
Co
2
% (start with 1%)
2. Let instrument run and wait for the
CO
2
% to stabilize
3. Measure
CO
2
% inside chamber
4. Input the difference between set
CO
2
% and actual CO
2
% with “ ” or “ ”
5. Repeat steps 1-4 for 3%, 5%, and 10%