background image

31

32

9. Operating Instructions

10. Maintenance

Instructions for Operation:

1. Only trained and professional personnel are allowed to operate the instrument.
2. It is forbidden to put flammable and explosive reagents into the instrument, and the
    instrument can not be immersed in the water.
3. Fully understand the nature of the chemical and the nature of the product created from 
    any chemical reactions to avoid any accidents or damage on the instrument. When 
    shaking volatile items, please do not close the door of instrument to prevent 
    explosions, high pressure, liquid spray, or other dangerous actions.
4. Place the load symmetrically. Uneven or unbalanced sample loading can cause 
    rocking of the instrument, creating a hazard.
5. When the heating or cooling is in use, make sure the sample does not affect the
    normal ventilation cycle inside the instrument.
6. Ensure the sample has been completely fixed before shaking, when restarting the 
    instrument check whether the sample needs to be re-fixed. When abnormal noise is 
    produced during shaking, reduce the speed or stop running then re-fix the sample 
    until the abnormality is eliminated.
7. Before running or stopping the instrument, confirm that the speed is set to minimum.
8. If the mechanical movement is unexpectedly interrupted then it may restart with no 
    warning.
9. Wear suitable protective equipment when operating the instrument to prevent damage 
    from liquid splashes, toxic or flammable gases exhausts.
10. After an unexpected interruption set the operating temperature to ambient 
      temperature to the maximum extent possible, avoid unnecessary burns, frostbite,
      or damage to the sample.
11. Do not block the vents to ensure smooth ventilation around the instrument.
12. When CO

2

 is connected, ensure the instruments CO

2

 connection port is tight and the 

      seal has no leakage.
13. Do not switch on/off too frequently.
14. Reference this user manual for safety warning symbol explanations.
15. Please comply with the user manual when using the instrument, otherwise it may 
      cause the instrument to malfunction or damage the instrument.
16. Please unplug before connecting water to the water inlet, do not flood the instrument.
17. High humidity can easily lead to decreased electrical insulation and the electrical 
      current leakage risk increases, pay attention to electrical safety when operating under 
      high humidity.
18. Please make sure the air circulation is sufficient when using CO

2

. The CO

2

%

  

will

      increase in a small confined room and can harm the operator’s health. Also, please 
      avoid direct inhalation of the CO

2

 gas when opening the incubator door.

19. No experiment samples are allowed in the electrical and mechanical inner workings,
      keep samples in the chamber.
20. Do not use excessive force when opening and closing the incubator door to prevent 
      the instrument from being damaged.

21. Please reduce the frequency of opening the incubator door to help maintain constant
      temperature and save CO

2

.

22. After the instrument’s cooling function is used for 10 days continuously, a heating
      de-humidification process is recommended. Recommendation: heat the instrument to
      45 °C for half an hour.

Maintenance Instructions: 

1. Do not let gasoline, oil, or any other volatile chemicals come into contact with the 
    outside coating of the instrument.
2. No experiment samples are allowed in the instrument’s inner workings.
3. Inspect the fastening screws on the flask clamp regularly to prevent them from dislodging.
4. Please clean the instrument regularly.
5. Do not clean the instrument with corrosive or flammable liquid and make sure these
    types of liquids are kept away from the instrument.
6. Please take off the side plates at the right and back side of the instrument to clean 
    out dust on the condenser and fan to ensure good heat dissipation and high cooling 
    efficiency.
7. Only use refrigerant that is consistent with what is noted on the nameplate (on left side 
    wall) of the instrument, and do not fill excessively.
8. The power supply must be unplugged when machine is under maintenance.
9. Use the specified fuse and make sure to disconnect power when replacing the fuse.
10. Regularly check the screws of rotation parts and connected parts to ensure they 
      are fastened and tight.  
11. Regularly check the light source of the instrument.
12. Regularly check the CO

2

 gas inlet.

 

 The refrigerant used in the instrument is flammable.

 

 Ensure good ventilation of the room.

 

 Do not use any mechanical means or other ways to accelerate the defrosting

   process.

 

 Do not use electrical appliances in the instrument except the ones our company

   recommends.

 

 It is forbidden to smoke, have an open flame, or use electrical appliances that 

   directly generate electrical sparks around the instrument.

 

 When repairing the compressor or refrigeration related pipeline, please move

    the instrument to an open space, do not maintain in the laboratory space.

   

 

Notice

Содержание AEOLUS IS-18A

Страница 1: ...AEOLUS Stackable Incubator Shaker...

Страница 2: ...odern biotechnology Our state of the art embedded microprocessor system utilized in this machine has powerful data processing ability outstanding stability excellent interference resistance and high p...

Страница 3: ...odern biotechnology Our state of the art embedded microprocessor system utilized in this machine has powerful data processing ability outstanding stability excellent interference resistance and high p...

Страница 4: ...antages of lightness sound insulation shockproof electrical insulation heat resistance cold resistance solvent resistance and aging resistance Automatic adjustment of PID control Instrument adjusts it...

Страница 5: ...antages of lightness sound insulation shockproof electrical insulation heat resistance cold resistance solvent resistance and aging resistance Automatic adjustment of PID control Instrument adjusts it...

Страница 6: ...am display and is easy to check Display can be switched from English interface to Chinese interface If user is used to the Chinese interface it is easy to switch at any time Remote WiFi monitor access...

Страница 7: ...am display and is easy to check Display can be switched from English interface to Chinese interface If user is used to the Chinese interface it is easy to switch at any time Remote WiFi monitor access...

Страница 8: ...x Humidity Door Open Direction Programmable Segments 12 Segments Relative Humidity Type Active Humidification Disinfection UV Lamps on Left and Right Walls Parameter Protection Screen Data Display Pas...

Страница 9: ...x Humidity Door Open Direction Programmable Segments 12 Segments Relative Humidity Type Active Humidification Disinfection UV Lamps on Left and Right Walls Parameter Protection Screen Data Display Pas...

Страница 10: ...nly Power Cord Liquid Collection Tray Fuse Cross Screwdriver 2 Single open end Wrench 24 Wrench S10 User Manual Filter Filter Protection Package Hood PU Pipe 5 m Silicone Tube 12 2 T Connectors 14 Two...

Страница 11: ...nly Power Cord Liquid Collection Tray Fuse Cross Screwdriver 2 Single open end Wrench 24 Wrench S10 User Manual Filter Filter Protection Package Hood PU Pipe 5 m Silicone Tube 12 2 T Connectors 14 Two...

Страница 12: ...The instrument must be placed on a firm and level surface The total weight of the instrument when fully loaded 474 lb unit must be taken into consideration There must be sufficient space left around t...

Страница 13: ...The instrument must be placed on a firm and level surface The total weight of the instrument when fully loaded 474 lb unit must be taken into consideration There must be sufficient space left around t...

Страница 14: ...eparation 1060mm 530mm 1290mm 880mm 340mm 530mm 530mm 1400mm 1290mm 530mm 530mm 530mm 1332mm 800mm 270mm 2055mm 80 mm Note When double and triple stacked it is recommended that the second and third in...

Страница 15: ...eparation 1060mm 530mm 1290mm 880mm 340mm 530mm 530mm 1400mm 1290mm 530mm 530mm 530mm 1332mm 800mm 270mm 2055mm 80 mm Note When double and triple stacked it is recommended that the second and third in...

Страница 16: ...he power box Changing the fuse Malfunctioning of the instrument Instrument is out of use for a long period of time 4 Safety Instructions 5 Installation Lifting the instrument by hand should be prohibi...

Страница 17: ...he power box Changing the fuse Malfunctioning of the instrument Instrument is out of use for a long period of time 4 Safety Instructions 5 Installation Lifting the instrument by hand should be prohibi...

Страница 18: ...filter first and then to the CO2 inlet located on the right side of the shaker See Figure 1 3 Set the CO2 secondary flow control valve to 0 08 0 1MPa 11 6 14 5psi WARNING Excessive pressure may cause...

Страница 19: ...filter first and then to the CO2 inlet located on the right side of the shaker See Figure 1 3 Set the CO2 secondary flow control valve to 0 08 0 1MPa 11 6 14 5psi WARNING Excessive pressure may cause...

Страница 20: ...d access port and tube and filter Note 5 Installation 17 18 Polyurethane Pipe Inner dia 7mm outer dia 10mm To the filter Stainless steel hoop To the CO2 regulator 6 Structure Down swing door Up swing...

Страница 21: ...d access port and tube and filter Note 5 Installation 17 18 Polyurethane Pipe Inner dia 7mm outer dia 10mm To the filter Stainless steel hoop To the CO2 regulator 6 Structure Down swing door Up swing...

Страница 22: ...e password needs to be input before some functions are modified The default password is 88888888 1 2 3 4 4 5 6 7 8 9 10 11 Main Interface 1 Displays the current status of the instrument 2 Light button...

Страница 23: ...e password needs to be input before some functions are modified The default password is 88888888 1 2 3 4 4 5 6 7 8 9 10 11 Main Interface 1 Displays the current status of the instrument 2 Light button...

Страница 24: ...3 Click the corresponding button to input the desired parameter value or click the corresponding ON OFF to control the instrument related function switch 4 After setting press Save and back or press...

Страница 25: ...3 Click the corresponding button to input the desired parameter value or click the corresponding ON OFF to control the instrument related function switch 4 After setting press Save and back or press...

Страница 26: ...address is valid when multi instrument networking 4 Pause when door is open Set the status to ON to keep the instrument running when the door is opened set the status to OFF to pause the instrument w...

Страница 27: ...address is valid when multi instrument networking 4 Pause when door is open Set the status to ON to keep the instrument running when the door is opened set the status to OFF to pause the instrument w...

Страница 28: ...dity inside chamber 4 Input the difference between set humidity and actual humidity with or 5 Repeat steps 1 4 for 50 70 and 90 0 0 0 0 0 0 0 0 1 0 3 0 5 0 10 0 c CO2 Calibrate CO2 Concentration can b...

Страница 29: ...dity inside chamber 4 Input the difference between set humidity and actual humidity with or 5 Repeat steps 1 4 for 50 70 and 90 0 0 0 0 0 0 0 0 1 0 3 0 5 0 10 0 c CO2 Calibrate CO2 Concentration can b...

Страница 30: ...CO2 The No Gas Alarm activates when the instrument does not receive a gas inject CO2 for over 5 minutes The main interface will show alarm information When an abnormality occurs in the instrument plea...

Страница 31: ...CO2 The No Gas Alarm activates when the instrument does not receive a gas inject CO2 for over 5 minutes The main interface will show alarm information When an abnormality occurs in the instrument plea...

Страница 32: ...ar Alarm History Clear all the historical alarm records saved by the instrument automatically 3 View Past Week s Data View the operation data of the instrument within the past week on the touchscreen...

Страница 33: ...ar Alarm History Clear all the historical alarm records saved by the instrument automatically 3 View Past Week s Data View the operation data of the instrument within the past week on the touchscreen...

Страница 34: ...ing under high humidity 18 Please make sure the air circulation is sufficient when using CO2 The CO2 will increase in a small confined room and can harm the operator s health Also please avoid direct...

Страница 35: ...ing under high humidity 18 Please make sure the air circulation is sufficient when using CO2 The CO2 will increase in a small confined room and can harm the operator s health Also please avoid direct...

Страница 36: ...is activated the instrument will continue to attempt to inject CO2 gas when the gas supply is resumed the alarm will be automatically cancelled Instructions Touchscreen off and Platform not shaking P...

Страница 37: ...is activated the instrument will continue to attempt to inject CO2 gas when the gas supply is resumed the alarm will be automatically cancelled Instructions Touchscreen off and Platform not shaking P...

Страница 38: ...om of the monitor iOS Android Scan QR code to download 12 2 Locking and Unlocking the Universal Aluminum Platform 1 When the instrument is not running unscrew the handle counterclockwise as in step 1...

Страница 39: ...om of the monitor iOS Android Scan QR code to download 12 2 Locking and Unlocking the Universal Aluminum Platform 1 When the instrument is not running unscrew the handle counterclockwise as in step 1...

Страница 40: ...s 13 Optional Accessories Model Name Specifications IS A50 IS A1 IS A2 IS A3 IS A4 IS A5 IS A6 IS A7 IS A8 IS A9 25ml 50ml 100ml 150ml 200ml 250ml 500ml 1000ml 2000ml 3000ml IS A11 IS A12 IS A23 IS A2...

Страница 41: ...s 13 Optional Accessories Model Name Specifications IS A50 IS A1 IS A2 IS A3 IS A4 IS A5 IS A6 IS A7 IS A8 IS A9 25ml 50ml 100ml 150ml 200ml 250ml 500ml 1000ml 2000ml 3000ml IS A11 IS A12 IS A23 IS A2...

Страница 42: ...antlement 4 No valid invoice except the goods can prove the validity of the warranty 5 Product model and serial number of products does not match the warranty certificate 6 The damage caused by force...

Страница 43: ...AEOLUS Stackable Incubator Shaker...

Отзывы: