8. Settings
8. Settings
5
.
0
0
.
0
0
.
0
0
.
0
0
.
0
16
.
0
28
.
0
37
.
0
b: Temperature Calibrate
a: Basic Set
2017-06-06 09:19:72 TUE
0
0
1
Language
choose
: ENGLISH
1.
Lock LCD Time
: Sets the time in minutes of inactivity after which the LCD screen
locks and the operator needs to input the password to unlock the instrument. Set
the Lock LCD Time to "0" to keep the LCD screen unlocked.
Close LCD Time
: Sets the time in minutes of inactivity after which the LCD screen
closes into its energy saver state (screen off). Set the Close LCD Time to "0" to
keep the LCD screen on.
2.
Time Calibrate
: Sets the display time, the new time will show in the upper right
corner of the main interface. Time is in 24-hour time.
3.
Communication Address
: Input a value from 1 to 32 (the communication address is
valid when multi-instrument networking).
4.
Pause when door is open
: Set the status to "ON" to keep the instrument running
when the door is opened; set the status to "OFF", to pause the instrument when the
door is opened and the alarm “Open door and paused the unit" is displayed.
5.
Light turns on when door is open
: Set the status to "ON" to turn the light on when the
door is open; set the status to "OFF" and the light will not turn on when the door is
open.
6.
WIFI:
Turn on when connected to IS-A64 Remote Monitoring accessory
7.
Language choose:
Choose between English and Chinese for the display language
Temperature can be calibrated by following the procedure below:
1. Set the instrument to the target temperature (start with 5
)
℃
2. Let instrument run and wait for the temperature to stabilize
3. Measure temperature inside chamber with thermometer
4. Input the difference between set temperature and actual temperature with “ ” or “ ”
5. Repeat steps 1-4 for 16
, 28
, and 37
℃
℃
℃