magnetic wand to engage the magnetic sensors that are on the side of the
enclosure. Refer to Section 4.1 Exterior Enclosure for location photo. This
is particularly useful if in a wet or dirty environment. A magnet of sufficient
strength will be required to trip the sensors. Such a magnet is included in
the Calibration Kit (see Section 8.4 Calibration Kit and can also be ordered
separately from CETCI under part number
CET-MW
.
To initiate non-intrusive calibration, touch the magnet to one of the marks
on the side of the enclosure and follow the screen prompts along with the
instructions in this section.
NOTE:
The programming buttons inside the enclosure can be used instead of
the magnetic wand, if preferred.
7.2 Calibrating the Internal Sensor(s)
NOTE:
If calibrating an Oxygen sensor, refer to Section 7.5 Calibrating an
Oxygen Sensor.
To calibrate the internal sensor(s), the user must go through the following
steps:
Step 1
In the Sensor Configuration Menu (passcode 3032), press either ARROW key
to scroll through the menu and select the channel you want calibrate. Press
ENTER and use the ARROW key to scroll to Calibrate Zero. Press ENTER and
use the ARROW keys to enter a value of 0.