LPT-M - Operation Manual
Press ENTER again to change the setpoint value by using the ARROW keys to
increase or decrease the number. Press ENTER to move to the next digit and
again to save and Exit the screen.
NOTE:
An Ascending alarm is used when the normal gas level starts out below
the alarm setting and when the reading goes above it, the alarm will trip. A
descending alarm is used when the normal gas level starts above the alarm
setting and when the reading goes below it, the alarm will trip. The descending
alarm is commonly used with Oxygen sensors.
6.11.2 MID Alarm Setpoint
Follow the same procedure as the
6.11.1 Setting the LOW Alarm Setpoint,
choosing the SP Mid menu option.
6.11.3 HIGH Alarm Setpoint
Follow the same procedure as the 6.11.1 Setting the LOW Alarm Setpoint,
choosing the SP High menu option.