Appendix A: LE220 (-plus and R-plus) Series Site or Transport Preparation
Placement of Equipment
The area required for the instrument (see drawing below) is approximately 41” wide by 30” deep by 21” high (104 cm x 76 cm x 53 cm). Leave
sufficient space beside the instrument and behind the instrument for cooling fan exhaust. Do not block rear panel vents or fans. The ultraCUBE may
be placed under the bench.
Power Requirements
Provide at least four standard outlets, one for the laptop computer, one for the LE220-plus, one for the CH05, and one for the WCS 2.0.
Your system draws a maximum of 12.5A. All components are configured with universal power supplies.
Note
: to avoid derating the CH05
cooling capacity, the CH05 may be used with a step-up transformer if operated at less than 115 VAC.
Water Requirements
The instrument’s water bath requires about 5.0 liters of either deionized or distilled water, which should be topped up daily. The water
should be changed and cleaning protocol with 10% bleach or equivalent solution performed at least once a month.
Relocating of the System
When moving the instrument within a laboratory, or between nearby laboratories, please employ the following guidelines:
•
Perform the System Shutdown process described earlier in this manual.
•
Close SonoLab and shut down the computer.
•
Disconnect power cords and USB cable. Disconnect the ultraCUBE.
•
Drain the water tank and reinstall in the instrument. Please contact Covaris or your local representative for assistance.
•
Upon arrival at the new location, follow the Setup and Installation instructions described earlier in this manual.
WARNING:
To prevent the possibility of electrical shock, always plug the system into a grounded power source.