LE220 (-plus, R-plus, and Rsc) Setup & Instruction Manual
22
Part Number: 010398 Rev H | November 2019
•
Motion Control Button
:
-
Home
: Moves all motion axes to their initialized position, as occurs when the software starts.
-
Load Position
: Moves the tray to the front of the safety enclosure to facilitate loading samples.
-
Start Position
: Takes the sample tray to the treatment position of the first row of the currently selected method. This position
may be used to verify that the water level in the bath is correct.
-
Service Position
: Raises the sample tray and the transducer assembly allowing the water bath to be removed from the
instrument.
-
Stop
: Stops any motion currently in progress.
3.4
Running a Method
When the “Run” button is green, the selected method may be started. Click the “
Run
” button to start. The instrument will
verify that water level is appropriate for the specified for the method, and begin the treatment.
Once a method starts, if the bath temperature rises beyond the user specified
limit, your system finishes the acoustic treatment of the current row of the
plate, then pauses, and automatically resumes treating the remaining rows
when the temperature is within limits, with no operator intervention required.
As the method runs, the graphic for the row under treatment flashes green,
and the actual average incident power and temperature data are displayed, as
shown.