Covaris
LE220 (-plus, R-plus, and Rsc) Setup & Instruction Manual
31
Part Number: 010398 Rev H | November 2019
-
Click the
+
button inside the process bar to display the step type menu.
-
Click on
Repeat
.
-
Click the
+
button inside the repeat bar to display the step type menu and create two or more treatment steps. See image below.
All listed treatment (or delay, etc.) steps will be applied to the first row selected for treatment, the number of times specified in the
Repeat bar. Then all will be applied to the second row, and so forth, until the entire plate is treated.
-
When all method parameters and steps have been defined, save changes.
In Scenario B, all treatment (or delay or prompt) steps are applied to all selected rows of the plate once. This repeats for the number of
times specified in the Repeat bar.
Scenario B is typically used in “pulsing”. That is, a row is treated with an acoustic dose, and is then allowed to rest while the remaining
rows are treated. Once the remaining rows are treated, a second acoustic dose is applied to each row.
Repeat these steps,
this many times